Determination of particle size distribution — Single particle light interaction methods — Part 4: Light scattering airborne particle counter for clean spaces

This document describes a calibration and verification method for a light scattering airborne particle counter (LSAPC), which is used to measure the size distribution and particle number concentration of particles suspended in air. The light scattering method described in this document is based on single particle measurements. The typical size range of particles measured by this method is between 0,1 μm and 10 μm in particle size.

Instruments that conform to this document are used for the classification of air cleanliness in cleanrooms and associated controlled environments in accordance with ISO 14644‑1 and ISO 14644‑2, as well as the measurement of number and size distribution of particles in various environments.

The following parameters are within the scope of this document:

— size setting error;

— counting efficiency;

— size resolution;

— false count;

— maximum particle number concentration;

— sampling flow rate error;

— sampling time error;

— response rate;

— calibration interval;

— reporting results from test and calibration.

Get this standard Prices exclude GST
PDF ( Single user document)
$256.00 NZD
HardCopy
$303.00 NZD
Networkable PDF
Price varies
Preview only close
Prev {{ page }}/ {{ numPages }} Next
Preview only close
Prev {{ page }}/ {{ numPages }} Next

Keep me up-to-date

Sign up to receive updates when there are changes to this standard

Related Information

Similar Standards

  • BS 1796-1:1989

    Test sieving, Methods using test sieves of woven wire cloth and perforated metal plate

  • BS 3406-1:1986

    Methods for determination of particle size distribution, Guide to powder sampling

  • BS 3406-4:1993

    Methods for determination of particle size distribution, Guide to microscope and image analysis methods

  • BS 3625:1963

    Specification for eyepiece and screen graticules for the determination of the particle size of powders

Preview only close
Prev {{ page }}/ {{ numPages }} Next
Preview only close
Prev {{ page }}/ {{ numPages }} Next

ISO 21501-4:2018

Get this standard Prices exclude GST
PDF ( Single user document)
$256.00 NZD
HardCopy
$303.00 NZD
Networkable PDF
Price varies

Request to add this standard to your subscription

ISO 21501-4:2018

Price varies
Online library subscription

Your organisation’s Account Administrator must approve a request to add a standard to your subscription.

You may add a comment to the administrator below.

Cancel